青木広宙, 古川亮, 西谷維心, 青山正人, 日浦慎作, 小南陽子, 松尾泰治, 吉田成人, 田中信治, 佐川立昌, 川崎洋
生体医工学 52(2) 97-107 2014年 査読有り
Calibration for calculating the external parameter of camera-projector system is important, to make high-precision 3-D measurement by 3-D endoscope using the grid active stereo. In this paper, we propose a new calibration method with a sphere calibration object in a geometrically flexible camera-projector system, and we examine the validity of the proposed method. As the result of comparing using conventional calibration with a plane calibration object, it is clarified that the proposed method resolves the scaling imprecision, that is a problem of the conventional method. Thus, the proposed method has higher potential of measurement accuracy. In addition, it is the advantage of the proposed method that precision estimation of calibration parameters with a low number of input images is possible.